Replacement low water sensor assembly for the Blue Myco B-310L, B-550L and B-1180L commercial sterilizers. Sight glass included.
The assembly mounts to the 1.5" fitting below and to the right of the manway. Three probes sit behind the sight glass and read the water level in the chamber. If those probes are not submerged, the sensor disables the controller and the machine will not run — and if the level drops below them mid-cycle, the unit shuts itself down.
That is deliberate. It is what stops the heating elements from firing dry, warping, and taking other components with them.
This is usually the part you need when:
- The unit will not power on even with water in the vessel
- The white HO-YW light stays dark
- Cycles shut down partway through for no apparent reason
- The sight glass is cracked, clouded or leaking
Before ordering, try cleaning the probes — sediment build-up is a common cause of false low-water readings and often fixes the fault outright. If the sight glass itself is damaged, stop using the cooker until it has been replaced.